2 Setting up the Bravo Platform
Starting up and shutting down
18 Bravo Platform User Guide
Shutting down the Bravo Platform
Shut down the Bravo Platform before you:
Clean the Bravo Platform
Change the liquid-handling head
Install accessories
Move the Bravo Platform
CAUTION
Always turn off the Bravo Platform before installing or uninstalling a liquid-handling
head. Failure to do so can damage the head electronics.
To shut down the Bravo Platform:
1 If removing the liquid-handling head, ensure that the disposable tips are removed.
2 Optionally, home the liquid-handling head.
3 Shut down the computer.
4 Turn off any accessories, for example, Pump Module.
5 If using an Auto Filling Reservoir, disconnect the bottles to prevent siphoning.
6 On the side of the Bravo Platform, press the power switch to the off (o) position.
7 If applicable, uninstall the liquid-handling head.
Related information
For information about... See...
Initializing the Bravo device “Initializing the device” on page 31
Turning on or off Bravo accessories “Bravo accessories overview” on page 104
Setting up the Bravo Platform “Workflow for setting up the Bravo
Platform” on page 16

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals