102  N9912A FieldFox User’s Guide  
Integration Bandwidth 
The Integration Bandwidth of the carrier and offsets is the frequency span over 
which power is measured. 
To change Integration Bandwidth: 
  Press Meas Setup 4 
  Then Integrating BW 
  Enter a value between 100 Hz and 100 MHz using the numeric keypad,  ▲|▼ 
arrows, or the rotary knob. 
  Press Enter 
How to define adjacent (Offset) channels 
An offset represents a range of frequencies over which power is measured. You 
can specify up to three offsets. The frequency range for each offset is specified 
with an Offset Freq and Integ BW.  Each offset that is created has a Lower 
(carrier MINUS Offset Freq) and Upper (carrier PLUS Offset Freq) set of 
frequencies.  
  Press Meas Setup 4 
  Then Offsets 
  Press Offset 1 2 3 to select an offset.  
  Then Offset State to set that offset ON or OFF. 
  Select Offset Freq This is the difference between the carrier center frequency 
and the center frequency of the offset channel.  Then enter a value using the 
numeric keypad.  
  Select Offset Integ BW This is the frequency range over which power is 
measured in that offset; half of the range below and half above the Offset Freq.  
How to set limits for adjacent channels 
You can set a unique threshold power for each of the offsets that will cause a 
FAIL indication (RED bar).  This occurs when the calculated dBc power (on top 
of the offset bar) is ABOVE the specified level.  
To set limits, with an ACPR measurement on the screen: 
  Press Limit   8  
  Then Set Limits (1,2, or 3) 
  Then choose from the following: 
  Lower Offset Sets a limit for the lower-frequency offset. 
  Upper Offset Sets a limit for the higher-frequency offset. 
  Then enter a value using the numeric keypad,  ▲|▼ arrows, or the rotary knob. 
  Then Limit Test ON OFF