SA (Spectrum Analyzer) Mode  99 
  Enter a value between 100 Hz and 3 GHz using the numeric keypad,  ▲|▼ 
arrows, or the rotary knob. 
  Press Enter 
Other Channel Power Settings 
All relevant FieldFox settings are made automatically to ensure the highest 
accuracy, such as ResBW, VideoBW, and sweep (SwpAcquisition) speed.  These, 
and all other  SA Mode settings, can be changed manually in a Channel Power 
measurement. 
  Learn about Measurement Preset on page 97. 
  Learn about Averaging on page 97. 
  Learn about RRC Weighting on page 103. 
Occupied Bandwidth 
Occupied Bandwidth measures the power of the displayed frequency span and 
displays vertical posts at the frequencies between which the specified percentage 
of the power is contained.  The frequency span between the two vertical posts is 
the Occupied Bandwidth. The Occupied Power, the power that is contained 
between the two posts, is also displayed in dBm.  
 
OBW measurement; Chan 190 Downlink; GSM850 Radio Standard 
How to select Occupied Bandwidth 
  Press Measure__1  
  Then Channel Measurements  
  Then Occupied BW 
When Occupied Bandwidth is selected, the following settings are maintained 
from a previous measurement: Center Frequency, Preamp ON|OFF, and RF 
Attenuation.