CALCulate:LIMit:CLEar[:IMMediate]
Clears front-panel indications of limits being exceeded and clears bit 11 ("Lower Limit Failed") and bit 12 ("Upper Limit
Failed") in the Condition Register of the Questionable Data Register event register group. The corresponding event
register bits are unaffected.
A condition register continuously monitors the state of the instrument. Condition register bits are updated in real time;
they are neither latched nor buffered.
An event register is a read-only register that latches events from the condition register. While an event bit is set,
subsequent events corresponding to that bit are ignored.
Parameter Typical Return
(none) (none)
Clear the limit test results:
CALC:LIM:CLE
l This command does not clear measurements in reading memory.
l The instrument clears front-panel indications of limits being exceeded and clears bits 11 and 12 in the
Questionable Data Register when the measurement function changes, or when any of the following commands is
executed:
CALCulate:LIMit:STATe ON
INITiate
MEASure:<function>?
READ?
CALCulate:LIMit:CLEar
*RST
SYSTem:PRESet
l To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
120 Agilent Truevolt Series DMM Operating and Service Guide