CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer}{<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer}?[{MIN|MAX|DEF}]
Sets the histogram's lower and upper range values. Setting the lower or upper range value
(CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer) disables automatic selection of the lower and upper range
values (CALCulate:TRANsform:HISTogram:RANGe:AUTO OFF).
Parameter Typical Return
-1.0E+15 to -1.0E-15,
or
0.0 (default)
or
+1.0E-15 to 1.0E+15
+1.00000000E+06
See Example.
l If automatic range value selection is enabled (CALCulate:TRANsform:HISTogram:RANGe:AUTO ON), the query
returns the computed range value. If no histogram data exists, 9.91E37 (Not a Number) is returned.
l Lower and upper range values are computed automatically if CALCulate:TRANsform:HISTogram:RANGe:AUTO is
ON.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
CALCulate:TRANsform:HISTogram[:STATe]{ON|1|OFF|0}
CALCulate:TRANsform:HISTogram[:STATe]?
Enables or disables histogram computation.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0ξ˜ƒ(OFF)ξ˜ƒorξ˜ƒ1ξ˜ƒ(ON)
See Example.
l The instrument turns this setting OFF when the measurement function is changed or after *RST or
SYSTem:PRESet.
Agilent Truevolt Series DMM Operating and Service Guide 125

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals