CALCulate:AVERage[:STATe]{ON|1|OFF|0}
CALCulate:AVERage[:STATe]?
Enables or disables statistics computation.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0ξ˜ƒ(OFF)ξ˜ƒorξ˜ƒ1ξ˜ƒ(ON)
Return the statistics of 100 frequency measurements:
CONF:FREQ 1.0E3,.001
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
*WAI
CALC:AVER:ALL?
Typical Response: +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01
l Statistics are cleared when the measurement function changes or when any of these commands is executed:
CALCulate:AVERage:STATe ON
CALCulate:AVERage:CLEar
INITiate
MEASure:<function>?
READ?
*RST
SYSTem:PRESet
l The instrument turns this setting OFF when the measurement function is changed or after *RST or
SYSTem:PRESet.