CALCulate:AVERage:CLEar[:IMMediate]
Clears all computed statistics: minimum, maximum, average, peak-to-peak, count and standard deviation.
Parameter Typical Return
(none) (none)
Clear the stored statistical data:
CALC:AVER:CLE
l This command does not clear measurements in reading memory.
l Statistics are cleared when the measurement function changes or when any of these commands is executed:
CALCulate:AVERage:STATe ON
CALCulate:AVERage:CLEar
INITiate
MEASure:<function>?
READ?
*RST
SYSTem:PRESet
l To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].