CALibration:COUNt?
Returns the calibration count. Read and record the initial count when you first receive your instrument.
Parameter Typical Return
(none) +117
Return the calibration count:
CAL:COUN?
l Because the value increments each time you store the calibration constants, a complete calibration adds many
counts. The calibration count also increments whenever you save a calibration string, change the calibration
password, or override the calibration security.
l You can perform this query regardless of whether the instrument is secured.
l This setting is non-volatile; it will not be changed by power cycling or *RST.
See Also
SYSTem:SECurity:COUNt?
CALibration:SECure:CODE<new_code>
Sets the security code to prevent unauthorized calibrations.
You must unsecure the instrument to calibrate it or to change the calibration string or security code.
Parameter Typical Return
Unquotedstringupto12characters
Must start with letter (A-Z)
May contain letters, numbers (0-9) and underscores
(none)
Set new security code:
CAL:SEC:STAT OFF,OLD_CAL_CODE
CAL:SEC:CODE TST_DUT165
CAL:SEC:STAT ON
l To change the code: unsecure calibration memory with the old code, then set the new code.
l If you forget your security code, see Unsecure Instrument Without Security Code.
l This setting is non-volatile; it will not be changed by power cycling or *RST.
134 Agilent Truevolt Series DMM Operating and Service Guide