SYSTem:TEMPerature?
Returns the instrument's internal temperature in °C.
Parameter Typical Return
(none) +2.85000000E+01
Return the instrument's internal temperature:
SYST:TEMP?
l The value returned is not affected by UNIT:TEMPerature.
SYSTem:TIME <hour>, <minute>, <second>
SYSTem:TIME?
Sets the time for the instrument's real-time clock. Use SYSTem:DATE to set the date.
Parameter Typical Return
<hour> 0 to 23
<minute> 0 to 59
<second> 0 to 60
20,15,30.000
Set instrument time to 20:15:30 (8:15:30 PM)
SYST:TIME 20,15,30
l The real-time clock is used for Mass Memory (MMEMory) system file timestamps.
l The real-time clock is battery-backed and retains the date and time when power is off.
l The date and time are not affected by *RST or SYSTem:PRESet.
l The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set the date
and time when you first receive your instrument.
SYSTem:UPTime
Returns the amount of time that the instrument has been running since the last power-on.
Parameter Typical Return
(none) +8,+2,+13,+50
Return time that the instrument has been running:
SYST:UPT
l This command is typically used to verify that the instrument is warmed up sufficiently before calibration.
l The numbers returned indicate days, hours, minutes, and seconsd, respectively.
SYSTem:VERSion?
Returns the version of SCPI (Standard Commands for Programmable Instruments) that the instrument complies with.
There is no front panel equivalent. See Introduction to the Introduction to the SCPI Language for details.
Parameter Typical Return
(none) 1994.0
Return the SCPI version:
SYST:VERS?
236 Agilent Truevolt Series DMM Operating and Service Guide