ML-2050A/2051A/2150A
3. Maintenance of the Laser Oscillator Section
(4) Remove the 2 screws and then remove the lead wire of the ashlamp. Remove
the 4 bolts and then remove the upper chamber. Put it upside down on a clean
cloth so that its inside may be visible.
The laser chamber consists of an upper part and a lower part. In the lower chamber,
the glass plate is put on the reector.
⇒
Put the removed reector and other parts on a clean cloth so that oil and dust
may not stick on them.
<Notes>
- Handle the laser chamber with extreme care to void attaching a aw or dust
onto the mirror face inside the reector. Such a aw or dust will reduce the
laser output.
- Do not touch the glass portion of the ashlamp directly with a hand or do not
blemish it. The ashlamp will be damaged. Before installing the ashlamp,
clean the glass portion of the lamp by using alcohol.
Upper chamber
Glass plate
Electrode cover
Remove the 2 screws.
Remove the 4 bolts.
Lead wire
Lead wire