Chapter 9: Shutting Down 151
10 Shut down the computer.
11 Wet a soft, lint-free cloth with DI water and gently wipe down the
following surfaces daily to remove salt buildup:
• SIT
Be careful not to bend the SIT.
• aspirator arm
Wipe all exposed surfaces of the aspirator arm that are outside the
cytometer chassis.
12 Use a dry lint-free cloth to remove residual moisture.
Tip Place an empty tube on the SIT to catch any BD shutdown solution
that might drip from the SIT.
13 Empty the condensation trap located underneath the fluidics cart power
panel. See Figure 1-6 on page 22.