48 BD FACSCanto II Instructions for Use
8 Lift the flow cell access door to check the flow cell for air bubbles.
Figure 4-2 Flow cell
• If you see bubbles choose Cytometer > Cleaning Modes > De-gas Flow
Cell.
• Click OK when the completion message appears.
• If you still see bubbles, repeat.
An error message might appear when you open the flow cell access door. To
dismiss the message, close the door and wait 30 seconds.
9 Check that laser warmup has finished.
check here for
bubbles
BD FACSCanto
BD FACSDiva