Chapter 5: Cytometer QC and Setup 73
Figure 5-6 Adjusting Threshold
3 (Optional) Enable and set another threshold parameter.
Select the associated checkbox.
Edit the threshold value.
Select Or (any of the parameter thresholds can be met) or And (all of
the parameter thresholds must be met).
Adjusting Spectral Overlap
Fluorochromes emit light over a range of wavelengths. During cytometer setup,
fluorescence spillover is automatically determined and corrected. If necessary,
you can use the spectral overlap controls to make manual adjustments:
1 Click in the Spillover % value field that needs correction (Figure 5-7 on
page 74).
2 Use one of the editing methods specified in Adjusting Detectors on page 71.
pointer