Chapter 6: Running Samples with BD FACSCanto Clinical Software 95
7 If no carousel is found, click to load tubes manually.
8 (Manual) When prompted, install the tube containing the first worklist
sample.
Push the aspirator arm to the left.
Place the tube on the SIT, ensure the tube is straight, and firmly push
up until the tube comes to a complete stop and is fully seated.
Center the aspirator arm under the tube. There are three sensor pins on
the aspirator arm. The bottom of the tube should sit within the center
of the pins.
9 Click .
Always wear gloves when manually loading samples. A fluid flush of the
exterior of the sample injection tube (SIT) occurs between samples that
might contain biohazardous waste.