5-22
Power-on test
The power-on test is passed. The power indicator and alarm system work
correctly, and the monitor starts up properly.
Performance test
ECG performance test and calibration
ECG waves are displayed correctly without noise and the displayed HR value is
within 80±1 bpm.
ECG Lead Off alarm behaves correctly.
Paced signals are detected and pace pulse marks are displayed when Paced is
set to Ye s .
The difference between the amplitude of the ECG calibration square wave and
that of the scale is not greater than 5%.
Resp test
The Resp wave is not distorted and the Resp value is within 40±2 rpm.
SpO
2
test
Measure SpO
2
on a healthy person's finger and a Pleth wave and PR value are
displayed. The displayed SpO
2
value is within 95%-100%.
NIBP tests and calibration
The difference is within ±3 mm when 0, 50 and 200 mmHg is set for NIBP
accuracy test.
There is no leakage with NIBP, or the manual leakage test result does not
exceed 6 mmHg/min.
Temp test
The value displayed for each Temp channel of the monitor is within 37±0.1°C.
IBP test and calibration
The static pressure value displayed for each IBP channel is within 200±4
mmHg.
The ART and LV waves for each IBP channel are displayed correctly.
Mainstream CO
2
module test
The mainstream CO
2
module is zeroed successfully and the waveform baseline
recovers to zero.
The CO
2
Apnea alarm behaves correctly.
The displayed CO
2
value is withing 45±2 mmHg.
Sidestream CO
2
module test and calibration
Block the gas inlet of the module or water trap. The sidestream CO
2
flow rate is
less than 10ml/min, and the alarm message "CO
2
Airway Occluded" is given.
This indicates that there is no leakage.
The displayed CO
2
value is within 6.0±0.3%.
Microstream CO
2
module test and calibration
Block the gas inlet of the module or water trap for about 30 seconds. The
alarm message "CO
2
Airway Occluded" is given. It indicates that there is no
leakage.
The displayed CO
2
value is within 6.0±0.3%.
PiCCO test
The TB value displayed on the monitor is within 37±0.2°C.