5-23
The C.O. measurement result is normally displayed.
The displayed static pressure value of pArt is within 200±4 mmHg.
The displayed static pressure value of pCVP is within 20±1 mmHg.
The waveforms of pArt and pCVP are displayed correctly.
Analog output test
The waves displayed on the oscilloscope are identical with those displayed on
the monitor.
Electrical safety tests
The enclosure leakage current is not greater than 100 µA in normal condition
and is not greater than 300 µA in single fault condition.
The earth leakage current is not greater than 300 µA in normal condition and
is not greater than 1000 µA in single fault condition.
The patient leakage current is not greater than 10 µA in normal condition and
is not greater than 50 µA in single fault condition.
The patient auxiliary current is not greater than 10 µA in normal condition and
is not greater than 50 µA in single fault condition.
Network print test
The network printer can print ECG reports correctly.
Battery check
When the AC power supply of the adapter is disconnected, the monitor can be
powered by the battery and operate properly.
When the AC power supply on the Dock is disconnected, the monitor can be
powered by the battery and operate properly.
The operating time of the battery meets the product specification.
Test conclusion:
Signature of tester: Date: