6-3
Message Severity Alarm Clearing Causes and Countermeasures
Error storing PDF
file
Prompt / The PDF file path settings on the
printer server and the PDFCreator
are not consistent or the PDF
storage space is full. Check the PDF
file path
settings for consistency, or delete
the files saved under the PDF file
path to release storage space.
Switch does not
support auto
retrieving bed No.
Low Cannot be cleared Replace the switch with another one
that supports CDP or LLDP.
Failed to auto
retrieve bed No.
information
Low Cannot be cleared Check whether the monitor is
connected with the Masterserver;
on the Masterserver, confirm that
the interface for monitor connection
is configured with bed No. binding.
Monitor Version
Low
Low Can be cleared The N Series system software
version is too low. In order for the N
Series monitor to connect to
the N1 which connects to the
external IBP module and perform
IBP monitoring, the system software
of N 1 and N series monitor should
be
V02.25 and above.
XX V overvoltage High Cannot be cleared Restart the machine. If the problem
persists after restart, the power may
be faulty. In this case, replace the
main control board or power supply
board.
XX V undervoltage High Cannot be cleared Restart the machine. If the problem
persists after restart, the power may
be faulty. In this case, replace the
main control board or power supply
board.
Battery
temperature too
high
High Cannot be cleared Check whether the ambient
temperature exceeds the allowable
operating range; if not, the battery
may be faulty. In this case, replace
the battery.
Incorrect number
of batteries
Low Cannot be cleared Check the number of batteries.
Transport module
battery is pulled
High Cannot be cleared Restart the machine. If the problem
persists after restart, the battery