6-4
Message Severity Alarm Clearing Causes and Countermeasures
out may be faulty. In this case, replace
the battery.
Transport module
battery is aged,
replace the battery
Low Audible and visual display
can be cleared
Replace the battery.
Error reading dock
E2PROM
High Cannot be cleared Re-plug N1/T1 (pay attention to
align the pins). If the problem
persists, accessing Factory
Maintenance Menu to format dock
EEPROM. Then, perform
crossmatching on dock, Rack and
N1/T1 main unit to locate the fault.
Power Board
Comm Error
High
Cannot be cleared Restart the monitor. If the alarm
persists, replace the power board or
the main control board.
Battery Error
High
Cannot be cleared Reinstall the battery first. If the
alarm persists, replace the battery. If
the alarm still persists, replace the
power board or the main control
board.
RT Clock Need
Reset
High
Cannot be cleared Rest the time, and restart the
monitor. If the alarm persists,
replace the button battery of the
main control board. If the alarm still
persists, replace the main control
board.
CMS/eGW
Disconnected
Low Audible and visual display
can be cleared
The monitor is disconnected from
the CMS. Check the network
connection.
Fail To Get WLAN
IP Address
Low Cannot be cleared Unable to automatically obtain the
wireless network IP address. Check
the network settings.
Fail To Get LAN1 IP
Address
Low Cannot be cleared Unable to automatically obtain the
wired network LAN1 IP address.
Check the network settings.
Loading Default
Config Failed
Low Can be cleared The default configuration is not
correctly loaded. The monitor will
restore to the factory default
configuration for the current patient
category.
The patient data
storage space is
nearly full. Please
delete some
Med Audible and visual display
can be cleared
Delete unnecessary earlier
discharged patient.