123
Auto Start Settings
The Auto Start Settings screen allows the operator to set parameters that will take effect
upon power up of the analyzer. The Auto Start Settings screen is accessed by pressing
from the Setup menu.
If the Auto Startup function is turned off, the analyzer will boot up with the same settings the
analyzer was last in. The operator can set the following parameters:
Auto Startup: Turn the Auto Startup function on or off.
Wait for (min): Allow for the time it takes for the analyzer to warm up before calibration.
Calibrations: Set the number of calibrations the analyzer will perform.
Start Range: Specify the range to set upon Power Up.
Access Level: Select the Operator Security Level.
Remote/Manual: Set whether the analyzer starts up in Remote or Manual mode.
THC/CH
4
Mode: Select the mode the analyzer starts up in; THC, CH
4
or THC/CH
4
/NMHC
mode.