EasyManua.ls Logo

CAI 700M - Verifying Span Deviations

CAI 700M
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
700M HFID 6-2-15
80
Verifying Span Deviations
Press from the Auto Cal Deviations menu to view the Verifying Span Deviations screen.
Measured Value is the averaged concentration during the Verifying Span stage of sequenced
and auto calibrations.
Variance is the difference of the measured value and span gas concentration.
% FS is the percent of full scale related to the calibrated range limit.