700M HFID 6-2-15
73
Analog Hold on Cal
Analog Hold on Cal will hold the analog outputs to the last measured value during
calibration. If Analog Hold on Cal is Off the analog outputs will be live. The existing setting
(Off) is shown at the top right of the menu. To access the Analog Hold on Cal menu, press
from the Calibration Setup menu.
From the Analog Hold On Cal menu, press to turn Analog Hold On, which will hold the
analog outputs at the last measured value. You will return to the Calibration Setup menu. The
Calibration Setup menu shows the current setting at the bottom-right corner of the screen after
the ellipsis (. . .).
Example: Analog Hold on Cal . . . On.
From the Analog Hold On Cal menu, press to turn Analog Hold Off. You will return to the
Calibration Setup menu. The Calibration Setup menu shows the current setting at the bottom-
right corner of the screen after the ellipsis (. . .).
Example: Analog Hold on Cal . . . Off.