EasyManua.ls Logo

CAI 700M - Auto Calibration;Check

CAI 700M
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
700M HFID 6-2-15
67
Auto Calibration/Check
Auto Calibration/Check lets the operator select whether the analyzer actually calibrates, or
performs a check of the calibration. To access the Auto Cal/Check menu, press from the
Calibration Setup menu. The current setting is shown on the upper right corner of the screen.
Press to set the analyzer parameter to Calibrate. The setting will be saved and the analyzer
will return to the Calibration Setup menu. The Calibration Setup menu shows the current setting
on the right side of the screen after the ellipsis (. . .).
Example: AutoCal/Check . . . Cal.
Press to set the analyzer parameter to Check. The setting will be saved and the analyzer
will return to the Calibration Setup menu. The Calibration Setup menu shows the current setting
on the right side of the screen after the ellipsis (. . .).
Example: AutoCal/Check . . . Check.

Table of Contents

Related product manuals