Initiate Sequenced Check
Initiate Sequenced Check is a useful tool for setting up Auto Calibration. From the Auto
Calibration Menu screen, pressing initiates a sequenced calibration check. Rather than
initiating a calibration, it checks the validity of your most current calibration.
A sequenced calibration check performs all of the steps of a sequenced calibration with the
exception of the zero and span calibrations. It does not set new offsets, gains or any alarms.