EasyManua.ls Logo

CAI 700M - Calibration Setup

CAI 700M
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
700M HFID 6-2-15
61
Calibration Setup
The Calibration Setup menu provides all the parameters necessary for completing a
successful calibration. To access the Calibration Setup menu, select from the Calibration
menu.
All parameters on the Calibration Setup menu apply to Automated Calibration. The following
also apply to Manual Calibration: Cal Pump/Valves, Auto Cal/Check, Deviation Limits and
Analog Hold on Cal. All settings should be verified for correct information before a manual or
automated calibration is attempted.
Please note that the Calibration Setup menu shows the current settings on the right side of the
screen after the ellipsis (…). Example: Cal Pump/Valves . . . Valves.

Table of Contents

Related product manuals