Auto Calibration Parameters
→
→
→
→
Auto Calibration Parameters allows the operator to select the range, mode and choose
between Zero and All calibration gases (both zero and span). To access the AutoCal
Parameter screen, press from the Calibration Setup menu.
To navigate between parameters, use the up or down arrow to move the highlight to
the field you intend to change. Press to open the field and change the parameter. Press
again to close the field after you have made your change.
The first changeable parameter is the Range to be calibrated. Press to open the field and
change the range. Then select a range (from 1 – 4) for calibration. To select all ranges, press 0.
Press to close the field.
The second parameter the operator can change is the Mode. The 700M HFID analyzer can
calibrate in either THC or CH
4
mode. Press to open the field and select the mode using the
up or down arrows . Press to close the field.