700M HFID 6-2-15
51
If the zero calibration is unsuccessful, check the following:
1. Make sure the correct gas was introduced into the analyzer.
2. Verify the Diagnostic Values while flowing gas.
3. Check Zero Gas Deviations under Manual Deviations:
→
→
→
→
→
4. Check Maximum Calibration Error under Calibration Setup:
→
→
→
→
→
After a successful Manual Zero Calibration, press the button to return to the Manual
Calibration menu.