When the concentration number has stabilized, press to set the span calibration. The
calibration should now be complete.
If the span calibration was successful, the screen will say ***Saved Current***. If the calibration
was unsuccessful, the screen will say Outside Deviation Limits. See the Manual Zero Calibration
section for examples of screens showing successful and unsuccessful calibrations.
If the span calibration is unsuccessful, check the following:
1. Make sure the correct gas was introduced into the analyzer.
2. Verify the Diagnostic Values while flowing gas.
3. Check Span Gas Deviations under Manual Deviations:
4. Check
Maximum Calibration Error under Calibration Setup:
After a successful Manual Zero and Span Calibration, the analyzer is ready for use.