A112102 PD Test Checking Kit (Option)
8-1
8. A112102 PD Test Checking Kit (Option)
This is an optional module. The A112102 is a PD Test Checking Kit that can be set to simulate
abnormal discharge of different levels when the Model 11210 Battery Cell Insulation Tester is
equipped with PD Tester Function Card (A112100) or PD Analyzer Function Card (A112101)
for testing. It can also perform routine check on the Model 11210 for the embedded partial
discharge detection function when other standard DUTs are in use.
8.1 Connection for PD Simulation
The A112102 front panel has 4 voltage input terminals that are composed of two equipotential
BNC terminals and two equipotential concave terminals. The lower connectors can use the
enclosed test cables to connect to the 11210 leakage current test terminals while the other
two connectors can connect to the DUT. Make sure that the DUT polarity is correctly
connected. The DUT negative terminal should connect to the A112102 HV(-) red concave
terminal, and the positive terminal should connect to the A112102 HV(+) BNC terminal as
shown in Figure 8-1.
Figure 8-1 Connection for PD Simulation
Be aware of high voltage output hazard as the voltage input terminal can
be up to DC 1000V. Do not touch the voltage input terminal, test cables
and DUT especially during testing to avoid electric shock. Do not remove
the DUT before the 11210 test procedures are completed as voltage may
exist on the DUT and cause danger.