Specification Verification Procedure
A-5
11210 test result of leakage current measurement:
LC
Range
Condition
Tolerance
( B ) - (C )
DMM(nA)
( B )
Test Value(nA)
( C )
20 nA
2V / 200MΩ
± (5%* Imeas_1+1) (nA)
Imeas_1
200 nA
2V / 20MΩ
± (1%* Imeas_2+2) (nA)
Imeas_2
LC
Range
Condition
Tolerance
( B ) - (C )
DMM(uA)
( B )
Test Value(uA)
( C )
2 uA
20V / 20MΩ
± (0.3%* Imeas_3+0.006) (uA)
Imeas_3
20 uA
20V / 2MΩ
± (0.3%* Imeas_4+0.06) (uA)
Imeas_4
200 uA
20V / 200kΩ
± (0.3%* Imeas_5+0.6) (uA)
Imeas_5
LC
Range
Condition
Tolerance
( B ) - (C )
DMM(mA)
( B )
Test Value(mA)
( C )
2 mA
20V / 20 kΩ
± (0.3%* Imeas_6+0.006) (mA)
Imeas_6
20 mA
20V / 2 kΩ
± (0.3%* Imeas_7+0.06) (mA)
Imeas_7