Description of Panel
4-3
(6) INTERLOCK Connector
It is mainly for the fixture or instrument with protection device used, these two ends need
to be short-circuited thus the test terminal can output voltage and current. Use the
short-circuit plate to short-circuit it if the single unit is utilized.
(7) RS-232 Connector
Standard RS-232C connector
(8) HANDLER Connector
I/O connector for test result output signal and external trigger
(9) USB Connector (B-Type)
Compatible with USBTMC standard
(10) LAN Connector
10/100 type LAN connector
(11) PD Tester / PD Analyzer Function Card Slot (Option)
It is the slot for interface card. Remove the two screws on the cover in advance and put
the optional interface card and next to lock the hand screw on the interface card tightly if
put PD Tester or PD Analyzer interface card is required.