Setting for Operation
5-25
BDV : Breakdown voltage test. Touch the [Slider] to enable or disable the function. Green:
Enable the test function, Grey: Disable the test function.
(1) Mode
There are two test modes, Seq. and Stair, and the default is Seq.
Seq. Mode: Tests every step from charge, test to discharge.
Stair Mode: Tests every step from charge to test but only performs discharge at the
last step.
(2) Start
It sets the start test voltage. The range is 1V1000V, and the factory default is1V.
(3) End
The end test voltage which is the same as the Test Voltage parameter in page 1 that
cannot be set.
(4) Rest Time
It sets the interval time. The range is 5ms9.999s, and the factory default is 5 ms.
(5) Step Count
It sets the step count in the range of 2~1001 steps, and the factory default is 2.
(6) Vent CONT (It does not support PD Analyzer for use.)
It sets to continue testing by maintaining the test voltage infinitely (∞) with no time
limit when the last step of test ends. Touch the [Slider] to enable or disable it. Green:
Enable continuous test, Grey: Disable continuous test.
Mode = Seq.
Mode = Stair