Setting for Operation
5-25
BDV : Breakdown voltage test. Touch the [Slider] to enable or disable the function. Green:
Enable the test function, Grey: Disable the test function.
(1) Mode
There are two test modes, Seq. and Stair, and the default is Seq.
Seq. Mode: Tests every step from charge, test to discharge.
Stair Mode: Tests every step from charge to test but only performs discharge at the
last step.
(2) Start
It sets the start test voltage. The range is 1V~1000V, and the factory default is1V.
(3) End
The end test voltage which is the same as the Test Voltage parameter in page 1 that
cannot be set.
(4) Rest Time
It sets the interval time. The range is 5ms~9.999s, and the factory default is 5 ms.
(5) Step Count
It sets the step count in the range of 2~1001 steps, and the factory default is 2.
(6) Vent CONT (It does not support PD Analyzer for use.)
It sets to continue testing by maintaining the test voltage infinitely (∞) with no time
limit when the last step of test ends. Touch the [Slider] to enable or disable it. Green:
Enable continuous test, Grey: Disable continuous test.