STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 157
7
SAFETY FUNCTIONS
Referred
Sec
tion
Table
55 - Error List
(continued)
Countermeasures
1. Contact Service office.
2. Contact Service office.
1. Contact Service office.
2. Connect the cable
securely.
3. Contact Service office.
4. When the TMS
function
is
not in use, set the
"TMS
Option
" to "DISABLE"
with
display
unit
iDT
-001
or
STP
-Link
.
Probable causes
1. Failure of the TMS heater.
2. Failure of the TMS unit.
1. Disconnection of the TMS
sensor.
2. Disconnection of the
external
TMS connection
cable.
3. Failure of the
external
TMS
connectio
n cable.
4. Incorrect setting of the
TMS function.
Occurrence
condition
TMS temperature
control error
(when the TMS
unit
detection
temperature
becomes lower
than the setting
value by 10
°C
(-18 ºF) or more.
Error is not
detected for a
certain time after
the
STP pump
starts.
TMS sensor is not
connected when
the state "TMS
Option
" is
"ENABLE"
.
Pump operation
Decelerat
e and
stop
Decelerat
e and
stop
(Error
message
)
TMS Lower
Temp
TMS
Sensor Lost
"FAILURE"
LED
Steady
red