STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 64
4
OPERATION
4.9 Parallel port input/output signal
The remote input/output signal connector "X2 REMOTE" is used for input/output remote signals (see
Figure 36). This connector is of D-Sub (37 pins, socket type) type. The screw for connector is M2.6.
Note: The connector (pin type) is not included. Procure control equipment at your company.
Note: It is recommended to use a remote cable with shield type, and connect both terminals to the
ground.
Figure 36 - X2 REMOTE connector
4.9.1 Input signal pins
Choose contact input signal or voltage input signal. When using contact input, refer to Table 18, Table
19, and Figure 37. When using voltage input, refer to Table 20, Table 21, and Figure 38. The input
signal pins function only when the input operation port is set to the parallel port.
Two abbreviations are used in Table 18, Table 19 and Figure 37:
COM: Common Pin
IN: Input Pin