STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 72
4
OPERATION
4.9.3 Rotation INHIBIT signal
When using rotation INHIBIT signal, short-circuit the "INHIBIT ENABLE IN" signal. Relations between
rotation INHIBIT signal input and pump operation state are shown in Table 23.
Signal input
Pump operation
After short-circuit of rotation INHIBIT input
signal (A), the START operation is
performed (B).
INHIBIT IN
Close
Open
START operation
The pump accelerates when the START
operation is performed (B).
After the START operation (A), the rotation
INHIBIT input signal is short-circuited (B).
INHIBIT IN
Close
Open
START operation
The pump does not accelerate when the
START operation is performed (A).
The pump accelerates when the rotation
INHIBIT input signal is short-circuited (B).
The rotation INHIBIT input signal is opened
(A) during acceleration or normal
operation.
INHIBIT IN
Close
Open
START operation
The pump decelerates and stops when the
rotation INHIBIT input signal is opened (A).
After the operation (A), the pump continues
deceleration and stops even by
short-circuiting the rotation INHIBIT input
signal (B). However, when the START signal
is input, the pump will accelerate.
Table 23 - Rotation INHIBIT signal input
A B
A B
A B