EasyManuals Logo

Edwards STP-iXA3306C User Manual

Edwards STP-iXA3306C
181 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 162
MAINTENANCE AND INSP
ECTION
8
8.4.3 Recommended maintenance intervals for main parts
Internal components of the STP pump will be damaged from gas load, heat, and corrosion when
used long-term. Deterioration or abrasion of the internal parts of the STP pump will cause
unexpected failures. When overhauling the STP pump, replacement of main parts are recommended.
Refer to the following list as recommended maintenance intervals for main parts.
Part name
Recommended maintenance intervals
*1
Corrosive gas
Inert gas or
vacuum pumping only
Rotor blades
3 years
5 years
Rotor shaft
3 years
5 years
Motor, sensor, electromagnet,
connector
5 years
7 years
Air-cooling fan
5 years
PCBA in the control unit
7 years
Accessories
TMS valve
3 years
TMS heater
3 years
Connection cable
5 years
*1: Maintenance intervals of any parts may be shortened depending on the condition of use.
The costs of replacing parts that need to be replaced when deterioration or abrasion is observed will
be at the customers charge, even though a period is not exceeding the recommended maintenance
intervals.
When overhaul of the STP pump is needed, contact Service office.