Promag 53 PROFIBUS DP/PA Technical data
Endress+Hauser 159
10.1.5 Power supply
Electrical connections
→ 46
Supply voltage
85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
Cable entries
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8 to 12 mm; 0.31 to 0.47 in)
• Sensor cable entry for armored cables M20 x 1.5 (9.5 to 16 mm; 0.37 to 0.63 in)
• Threads for cable entries, ½" NPT, G ½"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8 to 12 mm; 0.31 to 0.47 in)
• Sensor cable entry for armored cables M20 x 1.5 (9.5 to 16 mm; 0.37 to 0.63 in)
• Threads for cable entries, ½" NPT, G ½"
Cable specifications (remote version)
→ 54
Power consumption
AC: <15 VA (including sensor)
DC: <15 W (including sensor)
Switch-on current:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle
• EEPROM or T-DAT saves measuring system data if power supply fails
• S-DAT: exchangeable data storage device which stores the data of the sensor (nominal
diameter, serial number, calibration factor, zero point, etc.)
Potential matching
→ 62
10.1.6 Performance characteristics
Reference operating conditions
To DIN EN 29104 and VDI/VDE 2641:
• Fluid temperature: +28 ± 2 °C (+82 ± 4 °F)
• Ambient temperature: +22 ± 2 °C (+72 ± 4 °F)
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 × DN
• Outlet run > 5 × DN
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.