Promag 53 PROFIBUS DP/PA Installation
Endress+Hauser 45
3.4 Post-installation check
Perform the following checks after installing the measuring device in the pipe:
Device condition/specifications Notes
Is the device damaged (visual inspection)? –
Does the device correspond to specifications at the measuring point,
including process temperature and pressure, ambient temperature,
minimum fluid conductivity, measuring range, etc.?
→ 157
Installation Notes
Does the arrow on the sensor nameplate match the direction of flow
through the pipe?
–
Is the position of the measuring electrode plane correct? → 14
Is the position of the empty pipe detection electrode correct? → 14
Were all screws tightened to the specified tightening torques when the
sensor was installed?
→ 20
Were the correct seals installed (type, material, installation)? → 35
Are the measuring point number and labeling correct (visual inspection)? –
Process environment / process conditions Notes
Are the inlet and outlet runs respected? Inlet run 5 × DN
Outlet run 2 × DN
Is the measuring device protected against moisture and direct sunlight? –
Is the sensor adequately protected against vibration (attachment, support)? Acceleration up to 2 g in accordance
with IEC 600 68-2-6 → 161