Operation Promag 53 PROFIBUS DP/PA
66 Endress+Hauser
5Operation
5.1 Quick operation guide
The user has a number of options for configuring and commissioning the device:
1. Local display (option) → 67
The local display enables you to read all important variables directly at the measuring
point, configure device-specific parameters in the field and perform commissioning.
2. Configuration programs → 74
The configuration of profile and device-specific parameters is primarily done via the
PROFIBUS interface. You can obtain special configuration and operating programs from
various manufacturers for these purposes.
3. Jumpers/miniature switches for hardware settings
– PROFIBUS DP → 77
– PROFIBUS PA→ 82
You can make the following hardware settings using a jumper or miniature switches on
the I/O board:
‣ Address mode configuration (select software or hardware addressing)
‣ Device bus address configuration (for hardware addressing)
‣ Hardware write protection enabling/disabling
!
Note!
A description of the configuration of the current output (active/passive) and the relay
output (NC contact/NO contact) is contained in the "Hardware settings" section. →
80
a0001318
Fig. 42: Methods of operating PROFIBUS PA/DP
1 Local display for device operation in the field (option)
2A Configuration/operating programs (e.g. FieldCare) for operation via PROFIBUS DP/PA
2B Configuration/operating program for operating by means of the FXA193 service interface (e.g. FieldCare)
3 Jumper/miniature switches for hardware settings (write protection, device address, address mode)
2A
3
1
EscEsc
E
-
+
Esc
E
+
-
XXX.XXX.XX
EscEsc
E
-
+
FXA193
2B