EasyManua.ls Logo

Endress+Hauser PROline promag 53 - List of System Error Messages

Endress+Hauser PROline promag 53
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Promag 53 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 133
For the selection FAILSAFE_TYPE LAST GOOD VALUE (factory setting):
If a valid output value was available before the failure:
If no valid output value was available before the failure:
•For FAILSAFE_TYPE WRONG VALUE:
For status information, see the table in the following section.
9.2.2 List of system error messages
Quality code (HEX) Quality status Quality substatus Limits
0x44
0x45
0x46
UNCERTAIN Last usable value
OK
Low
High
Quality code (HEX) Quality status Quality substatus Limits
0x4C
0x4D
0x4E
UNCERTAIN Initial Value
OK
Low
High
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts 146)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits
Depicted on the local display:
S = System error
$
= Fault message (with an effect on the inputs and outputs)
! = Notice message (without any effect on the inputs and outputs)
No. # 0xx Hardware error
001 S: CRITICAL FAILURE
$
: # 001
0x0F BAD Device
Failure
Constant ROM/RAM
failure
Cause:
ROM/RAM error. Error when accessing the
program memory (ROM) or random access
memory (RAM) of the processor.
Remedy:
Replace the amplifier board.
011 S: AMP HW EEPROM
$
: # 011
0x0F BAD Device
Failure
Constant Amplifier
EEPROM
failure
Cause:
Amplifier with faulty EEPROM
Remedy:
Replace the amplifier board.

Table of Contents

Other manuals for Endress+Hauser PROline promag 53

Related product manuals