Promag 53 PROFIBUS DP/PA Technical data
Endress+Hauser 161
"
Caution!
Do not exceed the min. and max. temperatures for the lining of the measuring tube
( "Medium temperature range").
Note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic
regions.
• If both fluid and ambient temperatures are high, install the transmitter at a remote
location from the sensor ( "Medium temperature range").
Storage temperature
The storage temperature corresponds to the operating temperature range of the transmitter
and sensor.
"
Caution!
• The measuring device must be protected against direct sunlight during storage in order to
avoid unacceptably high surface temperatures.
• Choose a storage location where moisture does not collect in the measuring device. This
will help prevent fungus and bacteria infestation which can damage the lining.
Degree of protection
• Standard: IP 67 (NEMA 4X) for transmitter and sensor
• Optional: IP 68 (NEMA 6P) for remote version of Promag W and Promag P sensors
Shock and vibration resistance
Acceleration up to 2 g in accordance with IEC 600 68-2-6
(High-temperature version: no data available)
CIP cleaning
"
Caution!
The maximum fluid temperature permitted for the measuring device must not be exceeded.
CIP cleaning possible:
Promag E (110 °C / 230 °F), Promag H/P
CIP cleaning not possible:
Promag L/W
SIP cleaning
"
Caution!
The maximum fluid temperature permitted for the measuring device must not be exceeded.
SIP cleaning possible:
Promag H, Promag P (with PFA lining)
SIP cleaning not possible:
Promag E/L/W
Electromagnetic compatibility (EMC)
• As per IEC/EN 61326 and NAMUR Recommendation NE 21
• Emission: to limit value for industry EN 55011