Technical data Promag 53 PROFIBUS DP/PA
160 Endress+Hauser
Maximum measured error
Standard: ±0.2% o.r. ± 2 mm/s (o.r. = of reading)
!
Note!
Supply-voltage fluctuations have no effect within the specified range.
a0004456
Fig. 74: Max. measured error in % of reading
Repeatability
Standard: max. ±0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
10.1.7 Installation
Installation instructions
12
Inlet and outlet runs
Inlet run: typically 5 × DN
Outlet run: typically 2 × DN
Length of connecting cable
The permissible cable length L
max
for the remote version depends on the conductivity of
the medium 19.
A minimum conductivity of 20 μS/cm is required to measure demineralized water.
10.1.8 Environment
Ambient temperature range
Transmitter:
Standard: –20 to +60 °C (–4 to +140 °F)
Optional: –40 to +60 °C (–40 to +140 °F)
!
Note!
At ambient temperatures below –20 °C (–4 °F), the readability of the display may be
impaired.
Sensor:
Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)