Troubleshooting Promag 53 PROFIBUS DP/PA
134 Endress+Hauser
012 S: AMP SW EEPROM
$
: # 012
0x0F BAD Device
Failure
Constant Amplifier
EEPROM data
inconsistent
Cause:
Error when accessing data of the measuring
amplifier EEPROM
Remedy:
The EEPROM data blocks in which an error has
occurred are displayed in the
TROUBLESHOOTING function (No. 8047).
1. Press Enter to acknowledge the errors in
question; default values are automatically
inserted instead of the faulty parameter
values.
2. The measuring device must be restarted
after fault elimination.
Access:
SUPERVISION SYSTEM OPERATION
SYSTEM RESET ( RESTART)
031 S: SENSOR HW DAT
$
: # 031
0x10
0x11
0x12
BAD Sensor
Failure
O.K.
Low
High
S-DAT failure /
S-DAT not
inserted
Cause:
1. S-DAT is not correctly plugged into the
amplifier board (or is missing).
2. S-DAT is defective.
Remedy:
1. Check whether the S-DAT is correctly
plugged into the amplifier board.
2. Replace the S-DAT if it is defective.
Check whether the new replacement DAT
is compatible with the existing electronics.
Check the:
- Spare part set number
- Hardware revision code
3. Replace measuring electronics boards if
necessary.
4. Plug the S-DAT into the amplifier board.
032 S: SENSOR SW DAT
$
: # 032
0x10
0x11
0x12
BAD Sensor
Failure
O.K.
Low
High
S-DAT data
inconsistent
041 S: TRANSM. HW-DAT
$
: # 041
0x0F BAD Device
Failure
Constant T-DAT failure Cause:
1. T-DAT is not correctly plugged into the
amplifier board (or is missing).
2. T-DAT is defective.
Remedy:
1. Check whether the T-DAT is correctly
plugged into the amplifier board.
2. Replace the T-DAT if it is defective.
Check whether the new replacement DAT
is compatible with the existing electronics.
Check the:
- Spare part set number
- Hardware revision code
3. Replace measuring electronics boards if
necessary.
4. Plug the T-DAT into the amplifier board.
042 S: TRANSM. SW-DAT
$
: # 042
0x0F BAD Device
Failure
Constant T-DAT data
inconsistent
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts → 146)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits