Promag 53 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 135
061 S: HW F-CHIP
$
: # 061
0x0F BAD Device
Failure
Constant F-CHIP defect/not
plugged
Cause:
Transmitter F-Chip:
– F-Chip is defective
– F-Chip is not plugged into the I/O board or is
missing.
Remedy:
1. Replace the F-CHIP.
2. Plug the F-CHIP into the I/O board.
No. # 1xx Software error
101 S: GAIN ERROR AMP
$
: # 101
0x0F BAD Device
Failure
Constant Gain Error
Amplifier
Cause:
Gain deviation compared to reference gain is
greater than 2%.
Remedy:
Replace the amplifier board.
121 S: A/C COMPATIB.
!: # 121
0x0F BAD Device
Failure
Constant Amplifier and I/O
board only
partially
compatible
Cause:
Due to different software
versions, I/O board and amplifier board are only
partially compatible (possibly restricted
functionality).
!
Note!
– The indication on the display as notice
message appears only for 30 seconds (with
listing in "Previous system condition"
function).
– This condition can occur if only one
electronics board has been exchanged; the
extended software functionality is not
available. The previously existing software
functionality is still working and the
measurement possible.
Remedy:
Module with lower software version has either
to be updated by "FieldCare" with the required
software version or the module has to be
replaced.
No. # 2xx Error in DAT/no communication
205 S: LOAD T-DAT
!: # 205
0x0F BAD Device
Failure
Constant Save to T-DAT
failed
Cause:
Data backup (downloading) to T–DAT failed, or
error when accessing (uploading) the
calibration values stored in the T–DAT.
Remedy:
1. Check whether the T-DAT is correctly
plugged into the amplifier board.
2. Replace the T-DAT if it is defective.
Before replacing the DAT, check that the
new, replacement DAT is compatible with
the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring electronics boards if
necessary
206 S: SAVE T-DAT
!: # 206
0x0F BAD Device
Failure
Constant Restore from T-
DAT failed
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts → 146)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits