Proline Prosonic Flow P 500 HART Commissioning
Endress+Hauser 101
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Pressure The Liquid hydrocarbons
option is selected in the
Medium parameter and the
API gravity option, Density
option or Reference density
option is selected in the Sound
velocity calculation mode
parameter.
Enter the process pressure for
the installation.
0.8 to 110 bar 1.01325 bar
API table selection The Liquid hydrocarbons
option is selected in the
Medium parameter and the
API gravity option, Density
option or Reference density
option is selected in the Sound
velocity calculation mode
parameter.
Select the API reference
conditions (temperature and
pressure) that apply for the
reference density specified.
API table 5/6
API table 23/24
API table 53/54
API table 59/60
Other
API table 23/24
Alternative pressure value Enter an alternative user-
defined reference value for the
pressure.
0.8 to 110 bar 1.01325 bar
Alternative temperature value Enter an alternative user-
defined reference value for the
temperature.
–10 to 110 °C 29.5 °C
Viscosity The User-specific liquid
option is selected in Medium
parameter.
Enter medium viscosity at
installation temperature.
0.01 to 10 000 mm²/
s
1 mm²/s
Pipe material Select pipe material. Carbon steel
Ductile cast iron
Stainless steel
1.4301 (UNS
S30400)
1.4401 (UNS
S31600)
1.4550 (UNS
S34700)
Hastelloy C
PVC
PE
LDPE
HDPE
GRP
PVDF
PA
PP
PTFE
Pyrex glass
Asbestos cement
Copper
Unknown pipe
material
Stainless steel
Pipe sound velocity The Unknown pipe material
option is selected in the Pipe
material parameter.
Enter sound velocity of pipe
material.
800.0 to 3 800.0 m/s 3 120.0 m/s
Pipe dimensions Select if pipe dimensions are
defined by diameter or
circumference.
Diameter
Pipe circumference
Diameter
Pipe circumference The Pipe circumference
option is selected in the Pipe
dimensions parameter.
Define the pipe circumference. 30 to 62 800 mm 314.159 mm
Pipe outer diameter The Diameter option is
selected in Pipe dimensions
parameter.
Define the outer diameter of
the pipe.
0 to 20 000 mm 100 mm