Proline Prosonic Flow P 500 HART Commissioning
Endress+Hauser 109
Parameter Prerequisite Description User interface / 
Selection / User
entry
Factory setting
Lower range value output In Current span parameter
(→  108), one of the
following options is selected:
4...20 mA NE (3.8...20.5
mA)
4...20 mA US (3.9...20.8
mA)
4...20 mA (4...20.5 mA)
0...20 mA (0...20.5 mA)
Enter lower range value for the
measured value range.
Signed floating-point
number
Depends on country:
m³/h
ft³/h
Upper range value output In Current span parameter
(→  108), one of the
following options is selected:
4...20 mA NE (3.8...20.5
mA)
4...20 mA US (3.9...20.8
mA)
4...20 mA (4...20.5 mA)
0...20 mA (0...20.5 mA)
Enter upper range value for
the measured value range.
Signed floating-point
number
Depends on country
and nominal
diameter
Fixed current The Fixed current option is
selected in the Current span
parameter (→  108).
Defines the fixed output
current.
0 to 22.5 mA 22.5 mA
Damping current output A process variable is selected
in the Assign current output
parameter (→  108) and
one of the following options is
selected in the Current span
parameter (→  108):
4...20 mA NE (3.8...20.5
mA)
4...20 mA US (3.9...20.8
mA)
4...20 mA (4...20.5 mA)
0...20 mA (0...20.5 mA)
Enter time constant for output
damping (PT1 element).
Damping reduces the effect of
fluctuations in the measured
value on the output signal.
0.0 to 999.9 s 1.0 s
Failure behavior current output A process variable is selected
in the Assign current output
parameter (→  108) and
one of the following options is
selected in the Current span
parameter (→  108):
4...20 mA NE (3.8...20.5
mA)
4...20 mA US (3.9...20.8
mA)
4...20 mA (4...20.5 mA)
0...20 mA (0...20.5 mA)
Select output behavior in the
event of a device alarm.
Min.
Max.
Last valid value
Actual value
Fixed value
Max.
Failure current The Defined value option is
selected in the Failure mode
parameter.
Enter current output value in
alarm condition.
0 to 22.5 mA 22.5 mA
* Visibility depends on order options or device settings