Proline Prosonic Flow P 500 HART Mounting
Endress+Hauser 23
The sensors can be arranged in different ways:
• Mounting arrangement for measurement with 1 sensor set (1 measuring path):
• The sensors are located on opposite sides of the measuring pipe (offset at 180° ):
measurement with 1 or 3 traverses.
• The sensors are located on the same side of the measuring pipe: measurement with 2
or 4 traverses
• Mounting for measurement with 2 sensor sets
2)
(2 measuring paths):
• 1 sensor of each sensor set is located at the opposite side of the measuring pipe (offset
by 180°): measurement with 1 or 3 traverses
• The sensors are located on the same side of the measuring pipe: measurement with 2
or 4 traverses
The sensor sets are arranged on the measuring pipe, offset by 90°.
Using 5 MHz sensors
Here, the rails of the two sensor sets are always arranged at an angle of 180° to one
another for all measurements with 1, 2, 3 or 4 traverses. The sensor functions are
assigned in the two rails via the transmitter electronics unit depending on the selected
number of traverses. It is not necessary to swap the cables in the transmitter between
the channels.
Use of high-temperature sensors
The measuring paths for a high-temperature measurement are preferably mounted
with 1 traverse on the pipe. If 2 measuring paths are being used, the individual paths
are arranged so they are offset by 180° from each other (X-arrangement).
Single-path measurement (1 sensor set)
Vertical mounting
A0042013
9 1 traverse
A0042014
10 2 traverses
Horizontal mounting
A0044304
11 1 traverse
A0044305
12 2 traverses
2) Do not swap the sensors of the two sensor sets, as this can affect the measurement performance.