Commissioning Proline Prowirl R 200 HART
88 Endress+Hauser
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign frequency output The Frequency option is
selected in the Operating
mode parameter (→  85)
parameter.
Select process variable for
frequency output.
Off
Volume flow
Corrected volume
flow
Mass flow
Flow velocity
Temperature
Pressure
Calculated
saturated steam
pressure
*
Total mass flow
*
Energy flow
*
Heat flow
difference
*
Off
Minimum frequency value In the Operating mode
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→  88):
Volume flow
Corrected volume flow
Mass flow
Flow velocity
Temperature
Pressure
Calculated saturated steam
pressure
*
Total mass flow
*
Energy flow
*
Heat flow difference
*
Enter minimum frequency. 0 to 1 000 Hz 0 Hz
Maximum frequency value In the Operating mode
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→  88):
Volume flow
Corrected volume flow
Mass flow
Flow velocity
Temperature
Pressure
Calculated saturated steam
pressure
*
Total mass flow
*
Energy flow
*
Heat flow difference
*
Enter maximum frequency. 0 to 1 000 Hz 1 000 Hz