Proline Prowirl R 200 HART Commissioning
Endress+Hauser 87
Parameter Prerequisite Description Selection / User
entry
Factory setting
Failure mode In the Operating mode
parameter, the Pulse option is
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→  86):
Volume flow
Corrected volume flow
Mass flow
Total mass flow
*
Energy flow
*
Heat flow difference
*
Define output behavior in
alarm condition.
Actual value
No pulses
No pulses
Invert output signal Invert the output signal. No
Yes
No
* Visibility depends on order options or device settings
Configuring the frequency output
Navigation
"Setup" menu → Pulse/frequency/switch output
Pulse/frequency/switch output
Assign frequency output
→  88
Minimum frequency value
→  88
Maximum frequency value
→  88
Measuring value at minimum
frequency
→  89
Measuring value at maximum
frequency
→  89
Failure mode
→  89
Failure frequency
→  90
Invert output signal
→  90