ETAS ES910.3-A - Requirements When Setting up with Other Modules; Fig. 5-1 Free Space Around the ES910.3-A

ETAS ES910.3-A
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ES910.3-A - User’s Guide 43
ETAS Getting Started
Do not stand the ES910.3-A module on its top side.
Never place objects on top of the module.
Keep the ES910.3-A module away from sources of heat and make sure
the module is not standing in direct sunlight.
Adhere to the minimum distances of at least 10 centimeters on the sides
and above the module (see Fig. 5-1 on page 43).
Do not operate the ES910.3-A module in completely enclosed containers.
Fig. 5-1 Free Space around the ES910.3-A
5.2.2 Requirements when Setting up with Other Modules
You must adhere to important requirements when setting up and operating the ES910.3-A with
other modules to ensure sufficient ventilation and avoid heat concentration:
ETAS recommends not to stack or assemble ES910.3-A modules on top
ofeach other particularly if these modules are to be operated simultane-
ously.
Please adhere to the general notes on setting up and operating the
ES910.3-A in chapter 5.2.1 on page 42!
CAUTION!
Heat concentration in the module possible!
The electronics can be damaged due to overheating.
Do not cover ventilation slits when setting up, assembling or connect-
ing the ES910.3-A module.
Adhere to the minimum distances at the sides and above.
CAUTION!
Heat concentration on the top of the module.
The electronics can be damaged due to overheating.
Do not stack several ES910.3-As on top of each other.
When stacking the ES910.3-A with other ETAS modules, it always has
to be on top.
10 cm
10 cm
10 cm
CA N 1
LIN 1
CA N 2
LIN 2
ETK
TT
ON
TX
ER
°C
1
2
R
H
ES910.3
A