Test Setup - Test Configurator, Timer, and System
134 88000 Series
Modify Structure Button
Framing, available for EtherBERT test application, allows the selection
of the test framing type; otherwise the framing is set to Framed Layer 2.
See Network on page 264 for more information on frame format.
The framing is not configurable and set to Framed Layer 2 for
40GE/100GE dual port topology.
Framed Layer 1: Frame of x bytes that allows connection to any
interface that complies with 802.3 Ethernet PHY or DWDM fibre.
Available with rate up to 10G WAN.
Framed Layer 2: Frames of x
1
bytes without network layer (None)
that complies with IEEE 802a Ethernet II standard.
Framed Layer 3/4 (default): Frames of x
1
bytes with UDP (default)
or TCP network layer that complies with IEEE 802a Ethernet II
standard.
1. To set the frame length, see Frame Size on page 189 for EtherBERT (Ethernet and EoOTN) and page 293 for
RFC 2544.